|
Wire Detector |
Basics
There are basically three parts to the system. Detector, Driver,
and Computer. The detector is in the enclosure, the driver is to
the left, next to the computer, which is dedicated to the system
and will soon be networked.
The detector has a bias setting on top. It is set to "9" and should be
left alone; any change to the bias is now done in the software.
The driver has three buttons:
BIAS ... SOURCE
SELECTION ... ON/OFF
The power should be left on, or allowed to warm up for a half hour.
The source selector should always be set to Auto. The bias is the
switch you must turn off ANY time you need to touch the detector
or anything near it.
The computer runs windows NT. The software you need to run our
system is GADDS offline
GADDS Software
To collect an image, start the GADDS offline software, go
to [Collect] in the menubar, and select [Add...]. It will give you
a window to choose exposure time, etc.
To save an image, go to [File] and select [Print]. Then print to a file.
I choose grayscale .tif so that it translates to Macs. There is also
a translation program, but I have yet to see how, or if, it works.
|
Using the
Simple Collimator |
The monochromator works well with enough intensity that the straight
pipe shouldn't be needed. In case that statement is wrong, like for
a thick/dense sample, a simple set up is shown here to get full beam
intensity. You might want to add a Ni filter foil to clean up the Cu
K-beta.
Do's & Don'ts
Don't force any of the screws. Copper and Aluminum are very soft
compared to the steel screws.
Don't overtighten the "shutter pressure rod" (detail C).
It is there to hold the copper column against the manual shutter
disc (detail D),
which should seal flush to be safe during the alignment procedure.
The column stand only needs a bit of pressure to hold the shutter
open.
Be careful attaching a sample to the sample holder (detail B),
it is lead -very soft- and the hole is approx. 0.7 mm- don't dent
it or clog it, or you will have to make another.
Adjust the maximum intensity by changing the three setscrews near
the sample. Leave the other three near the source alone once they
are centered on the source.
Copper collimator is just a simple tube, it is set up at a distance
of 120 mm from source to sample.
I tried to make all the holes I tapped 1/4-20
thread. The exceptions are the adjuster thumbscrews (metric 1.5-10
mm) and the sample holder setscrew (?-?, 3/32 allen
hex.) The Collimator set screws use a 1/8 allen.
|
Using the
Monochromator |
Startup Alignment
The last person to use the detector should have left it in working
order, but I wouldn't bet on it. Regardless, check the placement
of the beamstop using a setting of 40 kV, 1 mA and the Zr foil over
the end of the sample holder and collect for ONE (1) second. If you
can't see any intensity upon collection, make sure that the shutter
is open and the bias is on, and if you still can't see any,
turn up the current to 5 mA. Repeat with 10 and 20 mA, then collect
for about 20 s. If at any time the beamstop is visibly centered on
the stop, move the stop by bending the arms (Sorry, that is the best
under the given circumstances) then begin again at 40 kV, 1 mA, and
1 s. This may sound tedious, but it isn't, these are collections
times well under a minute, and the alternative is a burnt out detector.
Once it is in line, there shouldn't be any reson to move it, so will
be good for you and the next guys untill someone changes something.
N.B. I tried to get the bed lined up as well as I
could withthe beam, so moving the detector and beamstop shouldn't put anything
too far out of alignment. And, yes,you still need to check the intensities
from the start: 40 kV, 1 mA, and 1 s.
Running
Once the beamstop is set, a real experiment can be done, including
a background airscatter run. in the software, start collecting an
image. We have preliminarily found that a visible pattern shows in
about 5 seconds, but statistically smooth will take more like an
hour.
Shutdown
This is easy. Make sure to do all these, and in order
- Bias Off
- Shutter closed
- current down
- voltage down
- Off lever (red)
- replace brass and aluminum plates CAREFULLY
- Clean up any mess
- Backup your files
Overall Alignment
Requires bypass of safety locks
This list may not be perfect, so THINK about every step you
take.
This is for when the monochromator was removed and needs to be
put back on. It took me half a day, so don't expect too much less
time.
Bias should already be Off
If the base plate needs to be put back to the 26° angle, remove
the detector from the dovetail rail
Remove the plate from the straight setup (just two small 6-32 screws)
Attach the monochromator to the source base with the 5mm-0.8 screws being
careful to get the shutter disk inside the square cutout.
Attach the base plate to the source post at 26° (1 screw, 6-32)
Make sure the collimator tube is clear of the monochromator
Put the lead shield (with the hole) over the shutter/monochromator gap
Make sure the lead glass fluorescent indicator is on the output hole
of the monochromator.
Tape up the mechanical switches at the
top of the doors
Close doors and turn on beam to 40 kV, 10 mA.
Open shutter
If you think the angle is still set correctly on the monochromator, skip
to the next step with the *, if not, continue
Turn thumbscrew to get graphite crystal to 0°
Set the output angle to zero (6° on the top scale, don't ask me.)
Turn the tiny setscrews on the miniature dovetail on the monochromator
to change the position of the crystal until the glow is a half circle
- the crystal is centered. Turn the thumbscrew to set the 13.53° crystal
angle
Turn the output hole to 27°
*Check the intensity by turning the thumbscrew at the back of the monochromator
a couple degrees and maximize the intensity. Readjust the output angle
as needed.
Close the shutter
Remove the protective lead glass indicator and bring the column up to
the output hole
Clamp the stand firmly and set the angle with the "shutter pressure rod" Don't
move it again if you can help it.
Adjust the three setscrews so that the copper tube is in line with the
output hole
Mount the Detector very carefully
Eyeball the angle to the Detector
Close doors and Remove tape from safety
circuit switches and set up for regular data collection
Hang the Zr foil over the end of the open end of the copper tube and
take a 40 kV, 1 mA, 1 s pattern with the detector to check for rough
alignment of tube and beam
Adjust the OTHER three setscrews to bring it in line. I put a post-it
note corner on the computer screen at the center. You'll have to move
the beamstop
Put the sample holder on, set screw on top, and Zr flim over the end
and take another pattern to check for alignment. If nothing is visible,
it is either the beamstop is well aligned, or more likely the sample
holder isn't.
Adjust the 3 set screws near the sample holder until the aperture is
correctly in the middle of the beam, as always, at 40 kV, 1 mA, 1 s with
the Zr foil.
Align the beamstop
Go to alignment sequence above
|
| References |
Manual for the Detector is in the Martin lab. Two users manuals, one
for the system, one for the software.
Manual for the monochromator (English version) is stored in the
Martin lab, and one copy should be with Anne Huber (2220 H. H. Dow)
Cullity, B. D., Elements of X-ray Diffraction, Addison Wesley,
Reading, Mass, 1978.
Vainstein, B. K., Diffraction of X-rays by Chain Molecules,
Elsevier, Amsterdam, 1966.
Gunier, A., Fournet, G., Small Angle Scattering of X-rays,
John Wiley & Sons, London, 1955.
Baumg, P. K., "Electr"J. Col Sci.,36 (1), 71-79,
(1971).
|