Hazard Information |
|
|---|---|
Wire Detector |
There are basically three parts to the system. Detector, Driver, and Computer. The detector is in the enclosure, the driver is to the left, next to the computer, which is dedicated to the system and will soon be networked. The driver has three buttons: BIAS ... SOURCE SELECTION ... ON/OFF The power should be left on, or allowed to warm up for a half hour. The source selector should always be set to Auto. The bias is the switch you must turn off ANY time you need to touch the detector or anything near it. The computer runs windows NT. The software you need to run our system is GADDS offline GADDS Software
To collect an image, start the GADDS offline software, go to [Collect] in the menubar, and select [Add...]. It will give you a window to choose exposure time, etc.
|
Using the |
The monochromator works well with enough intensity that the straight pipe shouldn't be needed. In case that statement is wrong, like for a thick/dense sample, a simple set up is shown here to get full beam intensity. You might want to add a Ni filter foil to clean up the Cu K-beta.
Don't force any of the screws. Copper and Aluminum are very soft compared to the steel screws.
Don't overtighten the "shutter pressure rod" (detail C). It is there to hold the copper column against the manual shutter disc (detail D), which should seal flush to be safe during the alignment procedure. The column stand only needs a bit of pressure to hold the shutter open. Be careful attaching a sample to the sample holder (detail B), it is lead -very soft- and the hole is approx. 0.7 mm- don't dent it or clog it, or you will have to make another. Adjust the maximum intensity by changing the three setscrews near the sample. Leave the other three near the source alone once they are centered on the source. Copper collimator is just a simple tube, it is set up at a distance of 120 mm from source to sample. I tried to make all the holes I tapped 1/4-20 thread. The exceptions are the adjuster thumbscrews (metric 1.5-10 mm) and the sample holder setscrew (?-?, 3/32 allen hex.) The Collimator set screws use a 1/8 allen.
|
Using the |
The last person to use the detector should have left it in working order, but I wouldn't bet on it. Regardless, check the placement of the beamstop using a setting of 40 kV, 1 mA and the Zr foil over the end of the sample holder and collect for ONE (1) second. If you can't see any intensity upon collection, make sure that the shutter is open and the bias is on, and if you still can't see any, turn up the current to 5 mA. Repeat with 10 and 20 mA, then collect for about 20 s. If at any time the beamstop is visibly centered on the stop, move the stop by bending the arms (Sorry, that is the best under the given circumstances) then begin again at 40 kV, 1 mA, and 1 s. This may sound tedious, but it isn't, these are collections times well under a minute, and the alternative is a burnt out detector. Once it is in line, there shouldn't be any reson to move it, so will be good for you and the next guys untill someone changes something.
Once the beamstop is set, a real experiment can be done, including a background airscatter run. in the software, start collecting an image. We have preliminarily found that a visible pattern shows in about 5 seconds, but statistically smooth will take more like an hour.
This is easy. Make sure to do all these, and in order
Requires bypass of safety locks This list may not be perfect, so THINK about every step you take. This is for when the monochromator was removed and needs to be put back on. It took me half a day, so don't expect too much less time. Bias should already be Off
If the base plate needs to be put back to the 26° angle, remove the detector from the dovetail rail |
Examples / Methods |
|
Common Problems |
|
References |
Manual for the monochromator (English version) is stored in the Martin lab, and one copy should be with Anne Huber (2220 H. H. Dow) Cullity, B. D., Elements of X-ray Diffraction, Addison Wesley, Reading, Mass, 1978. Vainstein, B. K., Diffraction of X-rays by Chain Molecules, Elsevier, Amsterdam, 1966. Gunier, A., Fournet, G., Small Angle Scattering of X-rays, John Wiley & Sons, London, 1955. Baumg, P. K., "Electr"J. Col Sci.,36 (1), 71-79, (1971).
|
Email Us: mrg_people@umich.edu
MAIN | RESEARCH | PEOPLE | EQUIPMENT | PUBLICATIONS